In general follow the sterilization instructions of the manufacturer.
Sterilization Methods:
1. Autoclaving
|
2. Hot-Air Sterilization
- When surgical instruments are hot-air sterilized, please take care to load and operate the sterilizers properly.
- To ensure safe sterilization, the temperature should not be below 180°C but should not exceed 200°C as this may cause structural changes leading to irreversible damage, especially as far as microsurgical instrument are concerned.
- Instruments with parts of rubber, plastic, or textile as well as plastic-coated instruments and handles for electrodes are not suitable for hot-air sterilization.
- The general use of lubricating agents should be omitted prior to hot air sterilization. Only oil the joints and ratchets of surgical instruments.
3. Gas Sterilization
- Gas sterilization should only be used when no other method is suitable i.e. it is not recommended for surgical instrument.
- Components of motor line, Optical systems of rigid endoscopes and Flexible endoscopes can be gas sterilized.
4. Cold Sterilization
It is not recommended for surgical instruments due to the following reasons:
- Most cold sterilization solutions render instruments sterile only after a 10 hour immersion. This prolonged chemical action can be more detrimental to surgical instruments than the usual 20 minute autoclave cycle, resulting in corrosion and dulling of sharp edges on knives, scissors, hooks, etc.
- The solutions are not changed as recommended, causing undesirable suspensions, as blood, saline, proteins, medications, detergents, soap, and debris, to circulate forward to new instruments.
- The instruments generally are not rinsed with deionized (softened) distilled water after being removed from these somewhat corrosive solutions (Cold Sterilization Solutions).
- With time and use these solutions become ever more corrosive.